MatMeas-DMS-2000-High--and-Low-Temperature-Dielectric-Impedance-Spectroscopy-System
MatMeas-DMS-2000-High--and-Low-Temperature-Dielectric-Impedance-Spectrometer-–-Vacuum-Insulation
MatMeas-DMS-2000-High--and-Low-Temperature-Dielectric-Impedance-Spectrometer
MatMeas-DMS-2000-Series-High--and-Low-Temperature-Dielectric-Impedance-Spectrometer
MatMeas-DMS-2000-High--and-Low-Temperature-Dielectric-Impedance-Spectroscopy-System
MatMeas-DMS-2000-High--and-Low-Temperature-Dielectric-Impedance-Spectrometer-–-Vacuum-Insulation
MatMeas-DMS-2000-High--and-Low-Temperature-Dielectric-Impedance-Spectrometer
MatMeas-DMS-2000-Series-High--and-Low-Temperature-Dielectric-Impedance-Spectrometer

MatMeas DMS2000 High& Low Temperature Dielectric Impedance Spectrometer

ASTM D150-compliant dielectric impedance system for precise measurements from -160°C to 450°C in one integrated platform.

  • Ultra-wide temperature range: -160°C to 450°C
  • Parallel-plate capacitor method (combine ASTM D150 design)
  • Vacuum-insulated chamber prevents frosting & false peaks near 0°C
  • Liquid nitrogen cooling + resistive heating for stable control (±1°C)
  • Real-time software monitoring with safety alarms & auto pressure relief
Products Categories
Get A Free Quote
Contact Form

MatMeas DMS2000 Dielectric Impedance Spectroscopy

The DMS2000 series is built for evaluating dielectric properties of materials under extreme high and low temperature environments. It uses the parallel-plate capacitor principle and is developed in accordance with ASTM D150. With the optional LNP-95 liquid nitrogen injection system, the platform supports continuous temperature sweeps from -160°C to 450°C, helping researchers obtain stable, repeatable dielectric impedance and temperature-spectrum data for diverse R&D needs.

Dielectric Impedance Spectrometer Technical Data Shee

Parameter Value
Temperature Range -160°C to 450°C
Temperature Control Accuracy ±1°C
Heating Ramp Rate 0–10°C/min (typical 3°C/min)
Heating/Cooling Method Resistive heating / Liquid nitrogen cooling
Frequency Range (Single Channel) 20 Hz – 30 MHz
Frequency Range (Five Channels) 20 Hz – 10 MHz
Recommended Sample Size Diameter < 20 mm; Thickness < 5 mm
Electrode Design Parallel-plate capacitor; upper semi-spherical electrode + lower flat electrode
Chamber Vacuum-insulated to prevent frosting and artifacts near 0°C
Power Supply 220–240 V, 50/60 Hz
Dimensions (L×W×H) 547 × 380 × 326 mm
Weight 25 kg
Warranty 1 year

Get a Quote for DMS2000

Share your sample type and temperature/frequency requirements—we will recommend the best configuration and testing workflow.

Dielectric Impedance Spectrometer Applications

  • Dielectric property evaluation of polymers, ceramics, and composites
  • Ferroelectric and insulating material characterization
  • Temperature-dependent impedance spectroscopy (research & industrial R&D)
  • Material screening and performance comparison under extreme temperatures

MatMeas DMS2000 Dielectric Impedance Spectrometer Video

More Details

1. Wide Temperature Range (-160°C to 450°C)
2. Vacuum Insulation for Reliable Low-Temperature Testing
3.  ASTM D150-Based Measurement Principle
4. Safety & Automation (Liquid Level Alarm, Pressure Relief)
5. Software Monitoring and Troubleshooting

Related Products

Related News

Scroll to Top

Get A Free Quote Now !

Contact Form
If you have any questions, please do not hesitate to contact us.